Features and Specifications

TECHNICAL SPECIFICATIONS
Monitoring | Display with 6 monitoring parameters | |
Rate, Inspiratory Pressure, MAP, PEEP, Tidal Volume, Minute Volume, and Plateau Pressure |
||
Other Display | Breath delivered indicator | |
Alarm indication | ||
Mode selection |
ALARMS
Operator-adjustable | High and Low Tidal Volume | |
Fixed Alarms | Low and High PEEP, Low and High Inspiratory Pressure, Rate, Mandatory ventilation in BPAP-S/T mode |
|
Special Alarms | Mains power lost, battery low, battery critically low, loss of internal voltage, loss of gas supply, hardware failure, power off while ventilating, Power On Self-Test, fault, loss of flowrate signal during ventilation | |
Speaker | Multitonal, High, Medium and Low Priority | |
Backup Buzzer | Single tone, Hardware fault |
VENTILATION MODES
Type | Mode | Description |
Pressure | PRVC | PRVC targets a set tidal volume at a set respiratory rate, but also responds to a patient’s inspiratory effort |
Pressure | PC-CMV | PC-CMV is a mandatory ventilation mode that delivers a set constant pressure at a constant rate to the patient |
Invasive and Noninvasive |
BPAP-S/T | BPAP-S/T is a bilevel pressure support mode that supports spontaneous breathing of the patient, and defaults to mandatory ventilation if the patient does not initiate a breath |
STANDARDS
Designed to these standards, certifications pending |
IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2007, ISO 80601-2-12:2011 + Cor.:2011, IEC 80601-2-80, ISO 5356, ISO 80601-2-74:2017, ISO 10993:2018 ISO 18562-1:2017, ISO 18562-2:2017, ISO 18562-3:2017 ISO 18562-4:2017, ISO 62304:2015 |
ELECTRICAL & PNEUMATIC SPECIFICATIONS
Input Voltage | 100 to 240 VAC, 50/60 Hz | |
Power Consumption | 80 W to 100 W 1.0/0.6 A full load | |
Backup Battery Time | Typical 1 hour, 20 minutes minimum | |
Oxygen Supply | 40 to 72 psi (276 kPa to 496 kPa) up to 100 L/min max | |
Air Supply | 40 to 72 psi (276 kPa to 496 kPa) up to 100 L/min max | |
Degree of Protection | IPX21 |
ENVIRONMENT
Temperature | Operating: 5ºC to 40ºC (41ºF to 104ºF) | |
Storage: -20ºC to 60ºC (-4ºF to 140ºF) | ||
Humidity | 10% to 95% non condensing (operating and storage) | |
Altitude | Up to approx. 2,000 m (6,562 ft) atmospheric pressure 600 to 1,100 hPa |
CONTROL SETTINGS & ADJUSTMENTS
Setting | Range | Increments |
PEEP | 5 to 25 (cmH2O) | 2.5 |
Inspiratory Pressure | 10 to 50 (cmH2O) | 5 |
Rate | 10 to 40 (BPM) | 2 |
I:E Ratio | 1:1 to 1:3 | 0.5 |
Sensitivity (Pressure Trigger) |
-3 to -0.5 (cmH2O) | 0.25 |
Tidal Volume | 100 to 800 (ml) | 50 |
Tidal Volume Alarm |
N/A | ±50% of setpoint or user adjustable in 50 ml increments |
FiO2 | 21% to 100% | 21, 30, 40, 50, 60, 70, 80, 90, 100% |
CONTROLS
PRVC Mode (Blue LED Indicators for active controls) |
PEEP Setpoint | |
Inspiratory Pressure Setpoint | ||
Rate | ||
I:E Ratio | ||
Sensitivity (Pressure Trigger) | ||
Tidal Volume Target | ||
PC-CMV (Green LED indicators for active controls) |
PEEP Setpoint | |
Inspiratory Pressure Setpoint | ||
Rate | ||
I:E Ratio | ||
Tidal Volume Alarm Limits (Default or User Set) | ||
BPAP-S/T (Amber LED indicators for active controls) |
PEEP Setpoint | |
Inspiratory Pressure Setpoint | ||
Rate | ||
I:E Ratio | ||
Sensitivity (Pressure Trigger) | ||
Tidal Volume Alarm Limits (Default or User Set) | ||
15 second Apneic Window (Default setting only) | ||
Physical Dimensions | Size | Length=328mm, Width=435mm, Height=243mm |
Weight | 15 lbs. | |
Display | 4 lines, 20 characters per line | |
Main patient outlet | 22mm | |
Oxygen inlet | DISS | |
Air inlet | DISS |