Features and Specifications

TECHNICAL SPECIFICATIONS
Monitoring   Display with 6 monitoring parameters
  Rate, Inspiratory Pressure, MAP, PEEP,
Tidal Volume, Minute Volume
Other Display Breath delivered indicator
Alarm indication
Mode selection
ALARMS
Operator-adjustable   High and Low Tidal Volume
Fixed Alarms   Low and High PEEP, Low and High Inspiratory Pressure,
Rate, Mandatory ventilation in BPAP-S/T mode
Special Alarms   Mains power lost, battery low, battery critically low, loss of internal voltage, loss of gas supply, hardware failure, power off while ventilating, Power On Self-Test, fault, loss of flowrate signal during ventilation
Speaker   Multitonal, High, Medium and Low Priority
Backup Buzzer Single tone, Hardware fault
VENTILATION MODES
Type Mode Description
Pressure PRVC PRVC targets a set tidal volume at a set respiratory rate,
but also responds to a patient’s inspiratory effort
Pressure PC-CMV PC-CMV is a mandatory ventilation mode that delivers
a set constant pressure at a constant rate to the patient
Invasive and
Noninvasive
BPAP-S/T BPAP-S/T is a bilevel pressure support mode that supports spontaneous breathing of the patient, and defaults to mandatory ventilation if the patient does not initiate a breath
VENTILATION MODES
Type Mode Description
Pressure PRVC PRVC targets a set tidal volume at a set respiratory rate,
but also responds to a patient’s inspiratory effort
Pressure PC-CMV PC-CMV is a mandatory ventilation mode that delivers
a set constant pressure at a constant rate to the patient
Invasive and
Noninvasive
BPAP-S/T BPAP-S/T is a bilevel pressure support mode that supports spontaneous breathing of the patient, and defaults to mandatory ventilation if the patient does not initiate a breath
STANDARDS
Designed to these standards,
certifications pending
IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2007,
ISO 80601-2-12:2011 + Cor.:2011, IEC 80601-2-80,
ISO 5356, ISO 80601-2-74:2017, ISO 10993:2018
ISO 18562-1:2017, ISO 18562-2:2017, ISO 18562-3:2017
ISO 18562-4:2017, ISO 62304:2015
ELECTRICAL & PNEUMATIC SPECIFICATIONS
Input Voltage   100 to 240 VAC, 50/60 Hz
Power Consumption   80 W to 100 W 1.0/0.6 A full load
Backup Battery Time   Typical 1 hour, 20 minutes minimum
Oxygen Supply   40 to 72 psi (276 kPa to 496 kPa) up to 100 L/min max
Air Supply   40 to 72 psi (276 kPa to 496 kPa) up to 100 L/min max
Degree of Protection   IPX21
ENVIRONMENT
Temperature Operating: 5ºC to 40ºC (41ºF to 104ºF)
  Storage: -20ºC to 60ºC (-4ºF to 140ºF)
Humidity   10% to 95% non condensing (operating and storage)
Altitude Up to approx. 2,000 m (6,562 ft)
atmospheric pressure 600 to 1,100 hPa
CONTROL SETTINGS & ADJUSTMENTS
Setting Range Increments
PEEP 5 to 25 (cmH2O) 2.5
Inspiratory Pressure 10 to 50 (cmH2O) 5
Rate 10 to 30 (BPM) 2
I:E Ratio 1:1 to 1:3 0.5
Sensitivity
(Pressure Trigger)
 -3 to -0.5 (cmH2O) 0.25
Tidal Volume 250 to 800 (ml) 50
Tidal Volume
Alarm
N/A ±50% of setpoint or user
adjustable in 50 ml increments
FiO2 21% to 100% 21, 30, 40, 50, 60, 70, 80, 90, 100%
CONTROLS
PRVC Mode
(Blue LED Indicators for active controls)
PEEP Setpoint
Inspiratory Pressure Setpoint
Rate
I:E Ratio
Sensitivity (Pressure Trigger)
Tidal Volume Target
PC-CMV
(Green LED indicators for active controls)
PEEP Setpoint
Inspiratory Pressure Setpoint
Rate
I:E Ratio
Tidal Volume Alarm Limits (Default or User Set)
BPAP-S/T
(Amber LED indicators for active controls)
PEEP Setpoint
Inspiratory Pressure Setpoint
Rate
I:E Ratio
Sensitivity (Pressure Trigger)
Tidal Volume Alarm Limits (Default or User Set)
15 second Apneic Window (Default setting only)
Physical Dimensions Size Length=328mm, Width=435mm, Height=243mm
Weight 15 lbs.
Display 4 lines, 20 characters per line
Main patient outlet 22mm
Oxygen inlet DISS
Air inlet DISS

BDR-19 Product/Spec Sheet